Comparison of noninvasive techniques to measure blood pressure in newborns
Manoel A S Ribeiro1, Humberto H Fiori, Jorge H Luz
1Pontifícia Universidade Católica do Rio Grande do Sul, Porto Alegre, RS, Brazil. anraquel@terra.com.br
Objective:
To compare blood pressure measurements in newborn infants using the flush method, pulse oximetry and oscillometry according to Doppler findings.
Methods:
Noninvasive blood pressure measurements were made using three techniques (flush method, pulse oximetry and oscillometry) in three groups of newborns: 15 normal term infants, 16 stable preterm infants, and 14 critically ill infants. All measurements were video recorded, edited separately, coded and analyzed independently by three neonatologists.
Results:
Fifty-seven measurements were made using each method. The flush method and pulse oximetry had a better correlation with Doppler findings than oscillometry (correlation coefficients: 0.89, 0.85, 0.71; p < 0.01). The difference between measurement means, their standard deviations and the 95% confidence intervals when compared with Doppler findings were: -5.2 ± 7.9 (-21.1:10.7) mmHg for the flush method; 0.4 ± 8.9 (-17.5:18.2) mmHg for pulse oximetry; and 6.4 ± 16.1 (-25.8:8.6) mmHg for oscillometry. The flush method had a better agreement with Doppler findings for the diagnosis of hypotension than oximetry and oscillometry.
Conclusions:
The flush method and pulse oximetry seem to be useful techniques to measure systolic blood pressure in newborn infants; oscillometry had the poorest agreement with Doppler findings to detect hypotension.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: