Related Experiment Video
Updated: Jun 5, 2026

Assessing Dyslexia at Six Year of Age
Published on: May 1, 2020
AAMI advice on Class 6 indicators
1SAVVY Sterilization Solutions for Healthcare, Woodbury, Minnesota, USA.
Abstract:
Class 6 emulating indicators may be used as internal chemical indicators and in PCDs to release nonimplant loads. Class 6 CIs could be used in BI PCDs for routine sterilizer efficacy monitoring and sterilizer qualification testing and along with BIs inside products undergoing product testing. In all applications, the Class 6 CIs/CI PCDs must be labeled for the specific cycle you are monitoring. Class 6 CIs and/or Class 6 CI PCDs are available from Getinge, Steris, SteriTec, and 3M.