Body weight-tailored contrast material injection protocol for 64-detector row computed tomography coronary
Tsuyoshi Isogai1, Masahiro Jinzaki, Yutaka Tanami
1Department of Diagnostic Radiology, Keio University School of Medicine, Tokyo, Japan.
Purpose:
The aim of this study was to assess the body weight-tailored contrast material injection protocol for 64-detector computed tomography (CT) to maintain optimal coronary CT attenuation.
Materials And Methods:
A total of 60 patients who underwent CT coronary angiography were assigned to one of three groups: protocol A, total of 0.7 ml kg(-1) with 300 mg I/ml iodine concentration; protocol B, total of 0.6 ml kg(-1) with 350 mg I/ml; protocol C, total of 0.7 ml kg(-1) with 350 mg I/ml. The injection duration was 10 s in all patients. The proportion of intracoronary attenuation values >325 HU was compared on a pervessel and per-patient basis.
Results:
The attenuation value was significantly more likely to exceed 325 HU with protocol C (95% for reader 1 and 90% for reader 2) than with protocol A or B on the per-vessel analysis (P < 0.05). The attenuation value also significantly exceeded 325 HU with protocol C (95% and 90%, respectively) more often than with protocol A or B on the per-patient analysis (P < 0.05).
Conclusion:
A total of at least 0.7 ml kg(-1) contrast material with 350 mg/ml iodine concentration is required to achieve optimal coronary CT attenuation (>325 HU) with 64-detector CT when using an injection duration of 10 s.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System IV: CMRI
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...