Related Experiment Video
Updated: Jun 4, 2026

Shrinkage of Dental Composite in Simulated Cavity Measured with Digital Image Correlation
Published on: July 21, 2014
Comparison of silorane and methacrylate-based composite resins on the curing light transmission
Ricardo Danil Guiraldo1, Simonides Consani, Rafael Leonardo Xediek Consani
1Department of Restorative Dentistry, School of Dentistry, University of North Paraná, Londrina, PR, Brazil. rdguiraldo@gmail.com
Abstract:
The aim of this study was to investigate the influence of different composite resins - Filtek P90 (silorane-based composite) and Heliomolar (methacrylate-based composite) - on light transmission and decrease in Knoop hardness between the bottom and top of cured specimens. The irradiance of a light-curing unit (LCU) was measured with a power meter (Ophir Optronics; 900 mw/cm2) and spectral distributions were obtained using a spectrometer (USB 2000). Twenty standardized cylindrical specimens (2 mm thick x 7 mm diameter) of each composite resin were obtained by curing using the LCU for 40 s. Light energy transmission through the composite was calculated (n=10). The Knoop hardness number for each surface was recorded as the mean of 3 indentations. The difference in Knoop hardness between the top and bottom (DKH) of the same specimen was calculated (n=10). The irradiance of light that passed through Filtek P90 (272 mW/cm2) was not significantly greater than that the passed through Heliomolar (271 mW/cm2). The DKH of Filtek P90 (25%) was significantly higher than that of Heliomolar (12%). There was a greater degree of subsurface polymerization of the methacrylate-based composite compared to the silorane-based composite.
