Measuring Maxillary Posterior Tooth Movement: A Model Assessment using Palatal and Dental Superimposition
Published on: February 23, 2024
Mechanical evaluation of space closure loops in orthodontics
Eduardo Uggeri Rodrigues1, Hiroshi Maruo, Odilon Guariza Filho
1Department of Orthodontics, Pontifical Catholic University of Paraná, Curitiba, Paraná, Brazil.
Abstract:
This study evaluated the mechanical performance of teardrop-shaped loops and teardrop-shaped loops with helix used in orthodontic space closure. Sixty retraction loops made with 0.019" x 0.025" stainless steel (SS) and beta-titanium (BT) wires were used. They were attached to a testing machine to measure the magnitudes of the sagittal force and the load-deflection ratio necessary for 1 mm, 2 mm and 3 mm activation. The results demonstrated that the BT alloy presented significantly smaller mean values (p < 0.01) of sagittal force and load-deflection than the SS alloy. The loop with the highest mean value of sagittal force and load-deflection was the teardrop-shaped loop (p < 0.01). Differences were observed in the mean values of sagittal force and load-deflection among activations, and the highest mean value was found in the activation of 3 mm, while the smallest mean value was evident in the activation of 1 mm (p < 0.01). It could be concluded that the metallic alloy used and the presence of a helix in configuration of the loops may have a strong influence on the sagittal force produced and on the load-deflection ratio; the teardrop-shaped loops and teardrop-shaped loops with helix in BT presented the release of lighter forces; the teardrop-shaped loop in SS generated a high load-deflection ratio, providing high magnitudes of horizontal force during its deactivation.