Systematic Hearing Performance Evaluation Process for Adolescents with Cochlear Implantation at Early Ages
Published on: March 24, 2023
Clinical evaluation of a new hearing aid anti-cardioid directivity pattern
H Gustav Mueller1, Jennifer Weber, Martina Bellanova
1Department of Hearing and Speech Sciences, Vanderbilt University, Nashville, USA. gus@gusmueller.net
Objective:
The purpose of this research was to evaluate a new directional hearing aid algorithm which automatically adapts to an anti-cardioid pattern in background noise when a speech signal originates from behind the hearing aid user.
Design:
Using the hearing-in-noise-test (HINT) in the soundfield, with the sentences delivered adaptively from the back (180°) and the standard HINT competing noise from the front (0°; 72 dB SPL), the participants were tested for three different hearing aid conditions: omnidirectional, conventional adaptive directional, and adaptive directional with the anti-cardioid algorithm enabled.
Study Sample:
Adults (n = 21) with bilaterally symmetrical downward sloping sensorineural hearing loss; experienced hearing aid users and aided bilaterally for experimental testing.
Results:
Results revealed a significant effect for the hearing aid microphone setting (p < .0001), with a HINT mean RTS of 4.2 dB for conventional adaptive directional, -0.1 dB for omnidirectional, and -5.7 dB when the anti-cardioid algorithm was active. This was a large effect size (Cohen's f2).
Conclusion: