Related Experiment Video
Updated: Jun 4, 2026

Esophageal Heat Transfer for Patient Temperature Control and Targeted Temperature Management
Published on: November 21, 2017
A new heat and moisture exchanger for laryngectomized patients: endotracheal temperature and humidity
Renske J Scheenstra1, Saar H Muller, Andrew Vincent
1Department of Head and Neck Oncology and Surgery, The Netherlands Cancer Institute-Antoni van Leeuwenhoek Hospital, Amsterdam, The Netherlands.
Objective:
To assess the endotracheal temperature and humidity and clinical effects of 2 models of a new heat and moisture exchanger (HME): Rplus, which has regular breathing resistance, and Lplus, which has lower breathing resistance.
Methods:
We measured endotracheal temperature and humidity in 10 laryngectomized patients, for 10 min each, with and without the HMEs. We sequentially tested 4 HME models (all Atos Medical, Hörby, Sweden), in randomized order: Rplus, Lplus, Provox Normal (the HME we regularly use and which we considered the reference HME), and Stomvent (an older HME model). We also assessed the short-term clinical and practical effects of the Rplus and Lplus in a prospective 3-week trial with 13 laryngectomized patients.
Results:
Rplus and Lplus had better humidification than Provox Normal (6.8 mg H(2)O/L, 4.3 mg H(2)O/L, and 3.7 mg H(2)O/L, respectively, P < .001), and no significant temperature difference. During the 3-week study period, 7 of the 13 patients reported noticeably lower mucus production with Rplus and Lplus.
Conclusions:
Rplus and Lplus had better heating and humidification than Provox Normal. Although Stomvent also performed well, its design is less convenient for laryngectomized patients. Further HME improvement is still warranted and should focus on improving the HME's heating capacity.
More Related Videos
09:34Thermal Preconditioning During Ex-vivo Lung Perfusion for the Rehabilitation of Damaged Lung Grafts before Transplantation
Published on: October 31, 2025
06:25Cooling or Warming the Esophagus to Reduce Esophageal Injury During Left Atrial Ablation in the Treatment of Atrial Fibrillation
Published on: March 15, 2020
Related Concept Videos
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...
Tracheostomy Care II: Procedure
Step 1: Perform hand hygiene, and put on personal protective equipment: gown, gloves, mask and...