Reliability of the M-FLEX™: equipment to measure palmar grasp strength in infants

Marcus Vinicius Marques Moraes1, Eloisa Tudella, Joyce Ribeiro

  • 1Department of Physiotherapy, Regional University of Blumenau, 105 Antonio da Veiga Street, Blumenau,SC, Brazil. moraes.mvm@gmail.com

Insights

This study validated the M-FLEX™ device for measuring infant palmar grasp strength. The instrument demonstrated reliable and reproducible results in infants from birth to four months old.

Area of Science:

  • Pediatric assessment
  • Infant motor development
  • Biomedical engineering

Background:

  • Palmar grasp is a key developmental milestone in infants.
  • Accurate measurement of palmar grasp strength is crucial for developmental assessments.
  • Existing methods may lack precision or standardization.

Purpose of the Study:

  • To determine the reliability and reproducibility of the M-FLEX™ device.
  • To assess the M-FLEX™ as an instrument for measuring palmar grasp strength in infants.
  • To characterize palmar grasp in infants aged 0-4 months.

Main Methods:

  • Thirty-nine healthy infants (0-4 months) were evaluated.
  • The M-FLEX™ device measured maximum grasp strength (FMAX), mean grasp strength (FMEAN), and grasp time (GRT).
  • Repeatability and reproducibility were assessed using ANOVA, Pearson's correlation, and ICC with 95% CI.

Main Results:

  • The M-FLEX™ device provided consistent and repeatable measurements of palmar grasp strength.
  • Reproducibility analysis across different sampling sites confirmed the device's reliability.
  • Statistical analysis (ANOVA, ICC) supported the instrument's validity.

Conclusions:

  • The M-FLEX™ is a reliable instrument for characterizing palmar grasp in infants.
  • This device can be a valuable tool in routine infant evaluations.
  • Further research can explore its application in clinical settings for developmental monitoring.