[Comparison of FBP and ML-EM reconstruction used by simulation data; effect of the projection number for the SPECT
Shuuji Fujihara1, Shingo Kageyama, Yasunori Isoda
1Department of Radiology, Matsue Red Cross Hospital, Japan.
Purpose:
The influence of the numbers of projection of SPECT exerting on a re-constructed image cannot be strictly evaluated by phantom studies. Therefore, we compared re-constructed images of the FBP method and the ML-EM method by using simulation data.
Methods:
Simulation data was entered in the image processing software, and the projection data that changed the numbers of projection was made. Afterwards, reconstructed images of the FBP and the ML-EM methods were compared with respect to contrast, %COV, and the NMSE value.
Result:
When the numbers of projection of the FBP and the ML-EM method were decreased, all of the contrast, %COV, and the NMSE value were more deteriorated than that of the ideal image. Therefore, the image quality of SPECT improves with both FBP and ML-EM methods when there are many numbers of projection. Moreover, the FBP method was excellent in a cold contrast, and the ML-EM method was uniformly excellent. Therefore, an understanding of features and their inspection are effective for the selection of each image reconstruction method.