Related Experiment Video
Updated: Jun 4, 2026

Transcutaneous Microcirculatory Imaging in Preterm Neonates
Published on: December 31, 2015
Accuracy of temporal artery thermometry in neonatal intensive care infants
Gail Lee1, Diane Flannery-Bergey, Kris Randall-Rollins
1Exempla Lutheran Medical Center, Wheat Ridge, Colorado 80033, USA. schroeders@exempla.org
Purpose:
To determine the accuracy of temporal artery and axillary temperatures and the discomfort level of stable neonates during temperature measurement.
Subjects:
Convenience sample of neonates between the ages of 32 and 40 weeks' gestation cared for in an isolette or crib.
Design:
A method-comparison design was used to compare different methods for noninvasive temperature monitoring (infra-red temporal artery; axillary electronic) to core body temperatures (indwelling rectal probe).
Main Outcome Measure:
Bias and precision of test temperature devices (temporal artery; axillary).
Results:
Bias and precision for the temporal artery and axillary devices were 0.30 ± 0.44 and 0.28 ± 0.33, respectively. Analysis of variance found significant differences between both temporal and axillary temperatures compared to rectal temperatures (P < .01). Statistical differences were small and did not represent a clinically important difference. No statistical difference was found between temporal artery and axillary temperatures (P = .81). Increases in neonate discomfort after temperature measurement were significantly greater with axillary than increases after temporal artery temperature measurement (P = .03).
Conclusions:
This study found that body temperature measured with the temporal artery thermometer was similar to temperatures obtained with an axillary thermometer in stable, afebrile neonates. The use of temporal artery thermometry appears to be an acceptable approach for noninvasive temperature measurement in neonates, which causes less discomfort in neonates.
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...