Related Experiment Video
Updated: Jun 4, 2026

06:20
Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
A practical approach to diode based in vivo dosimetry for intensity modulated radiotherapy
Nils Kadesjö1, Tufve Nyholm, Jörgen Olofsson
1Department of Radiation Sciences, Radiation Physics, Umeå University, Östra Norrlandsgatan 62B, Umeå, Sweden. kadelake@gmail.com
Abstract:
A method for in vivo entrance dosimetry, using point detectors, in intensity modulated radiotherapy has been clinically evaluated. Diode dosimetry was performed for treatments of the head and neck and prostate regions. The results were good; 92.2% of the measurements showed deviations within ±5% of the expected values.