Percutaneous excisional biopsy of clinically benign breast lesions with vacuum-assisted system: comparison of three
Zhi Li Wang1, Gang Liu, Yan Huang
1Department of Ultrasound, Clinical Division of Iatrotechnique, Chinese People's Liberation Army General Hospital, 28 Fuxing Road, Beijing 100853, China. wzllg@sina.com
Purpose:
The aim of this study was to compare three devices in percutaneous excisional biopsy of clinically benign breast lesions in terms of complete excision rate, duration of procedure and complications.
Materials And Methods:
In a retrospective study from March 2005 to May 2009, 983 lesions underwent ultrasound-guided excisional biopsy with three vacuum-assisted systems, respectively. The lesions were category 3 lesions as determined by ultrasound imaging according to Breast Imaging Reporting and Data System (BI-RADS) (n=951) or had been confirmed as benign by a previous core needle biopsy (n=32). The completely excision rate, duration of procedure and complications (hematoma, pain and ecchymosis) were recorded.
Results:
99.7% (980/983) lesions were demonstrated to be benign by pathology after percutaneous excisional biopsy. The overall complete excision rate was 94.8% (932/983). In lesions whose largest diameter equal to or larger than 1.5cm, the complete excision rates of EnCor(®) group (97.8%, 348/356) and Mammotome(®) group (97.2%, 139/143) were significantly higher than that of Vacora(®) group (91.9%, 445/484) (P<0.05). The EnCor(®) group (6.6±6.5min) had a significant less duration than Mammotome(®) (10.6±9.3min) and Vacora(®) group (25.6±23.3min) (P<0.05). Hematoma occurred more in EnCor(®) group and Mammotome(®) group than in Vacora(®) group (P<0.05).
Conclusions:
All these three vacuum-assisted systems are highly successful for excisional biopsy of benign breast lesions.