Related Experiment Video
Updated: Jun 4, 2026

Open Tracheostomy Gastric Acid Aspiration Murine Model of Acute Lung Injury Results in Maximal Acute Nonlethal Lung Injury
Published on: February 26, 2017
Massive aspiration past the tracheal tube cuff caused by closed tracheal suction system
Mital H Dave1, Angela Frotzler2, Caveh Madjdpour2
1Department of Anaesthesia, University Children's Hospital, Zurich, Switzerland. mital.dave@kispi.uzh.ch
Background:
Aspiration past the tracheal tube cuff has been recognized to be a risk factor for the development of ventilator-associated pneumonia (VAP). This study investigated the effect of closed tracheal suctioning on aspiration of fluid past the tracheal tube cuff in an in vitro benchtop model.
Methods:
High-volume low pressure tube cuffs of 7.5 mm internal diameter (ID) were placed in a 22 mm ID artificial trachea connected to a test lung. Positive pressure ventilation (PPV) with 15 cm H₂O peak inspiratory pressure and 5 cm H₂O positive end-expiratory pressure (PEEP) was used. A closed tracheal suction system (CTSS) catheter (size 14Fr) was attached to the tracheal tube and suction was performed for 5, 10, 15, or 20 seconds under 200 or 300 cm H₂O suction pressures. Amount of fluid (mL) aspirated along the tube cuff and the airway pressure changes were recorded for each suction procedure. Fluid aspiration during different suction conditions was compared using Kruskal-Wallis and Mann-Whitney test (Bonferroni correction [α = .01]).
Results:
During 10, 15, and 20 seconds suction, airway pressure consistently dropped down to -8 to -13 cm H₂O (P < .001) from the preset level. Fluid aspiration was never observed under PPV + PEEP but occurred always during suctioning. Aspiration along the tube cuff was higher with -300 cm H₂O than with -200 cm H₂O suction pressure (P < .001) and was much more during 15 and 20 seconds suction time as compared to 5 seconds (P < .001).
Conclusion:
Massive aspiration of fluid occurs along the tracheal tube cuff during suction with the closed tracheal suction system.
Related Concept Videos
Tracheostomy Suctioning I: Pre-Procedural Steps
Equipment Required
First, gather all necessary equipment: a sterile suction catheter, a sterile disposable container, sterile gloves, a towel or...
Tracheostomy Suctioning II: Procedure
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Trachea
Anatomical Features:
Location: About half of the trachea is situated in the neck, anterior to the esophagus, and extends from the larynx (at the level of the...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques