The Quantification of Injectability by Mechanical Testing
Published on: May 13, 2020
Injection force of SoloSTAR® compared with other disposable insulin pen devices at constant volume flow rates
1Sanofi-Aventis Deutschland GmbH, Site Frankfurt Devices/Device Industrialisation, Frankfurt, Germany. thomas.vanderburg@sanofi-aventis.com
Background:
Injection force is a particularly important practical aspect of therapy for patients with diabetes, especially those who have dexterity problems. This laboratory-based study compared the injection force of the SoloSTAR® insulin pen (SoloSTAR; sanofi-aventis) versus other available disposable pens at injection speeds based on the delivered volume of insulin released at the needle.
Method:
Four different prefilled disposable pens were tested: SoloSTAR containing insulin glargine; FlexPen® and the Next Generation FlexPen® (NGFP) (Novo Nordisk), both containing insulin detemir; and KwikPen® containing insulin lispro (Eli Lilly). All pens were investigated using the maximum dispense volume for each pen type [80 units (U) for SoloSTAR; 60 U for the other pens], from the free needle tip dispensing into a beaker. Twenty pens of each type were fitted with the recommended needles and tested at two dose speeds (6 and 10 U/s); each pen was tested twice.
Results:
Mean plateau injection force and maximum injection force were consistently lower with SoloSTAR compared with FlexPen, NGFP, and KwikPen at both injection speeds tested. An injection speed of 10 U/s was associated with higher injection force compared with 6 U/s for all the pens tested (p < .001).
Conclusions:
SoloSTAR stands out because of its low injection force, even when compared with newer insulin pen devices such as the KwikPen and NGFP. This may enable patients, especially those with dexterity problems, to administer insulin more easily and improve management of their diabetes.
Related Concept Videos
Insulin Formulations: Types and Delivery
Short-acting insulins are divided into rapid-acting...
Parenteral Drug Delivery Systems: Injectables, Implants, and Infusion Devices
One-Compartment Open Model for IV Bolus Administration: Estimation of Elimination Rate Constant, Half-Life and Volume of Distribution
Modified-Release Drug Delivery Systems: Rate-Programmed I
Modified-Release Drug Delivery Systems: Rate-Programmed II
One-Compartment Open Model for IV Bolus Administration: Estimation of Clearance
In the one-compartment open model for intravenous (IV) bolus administration, clearance is estimated by dividing the elimination rate by the plasma drug concentration. This equation leverages the elimination rate constant and the apparent...