Identifying Coronary Artery Calcification on Non-gated Computed Tomography Scans
Published on: August 28, 2018
Evaluation of CT coronary artery angiography with 320-row detector CT in a high-risk population
1Department of Medical Imaging, Jinan Military General Hospital, Jinan, Shandong Province, China. cjr.sungang@vip.163.com
Insights
320-detector row dynamic volume CT (DVCT) accurately detects coronary artery disease (CAD) without heart rate control. Controlling heart rate below 70 bpm significantly reduces radiation dose, making DVCT a valuable tool for high-risk patients.
Area of Science:
- Cardiovascular Imaging
- Radiology
- Medical Diagnostics
Background:
- Coronary artery disease (CAD) poses a significant health risk, necessitating accurate diagnostic methods.
- 320-detector row dynamic volume CT (DVCT) is an advanced imaging technique for evaluating coronary arteries.
- High-risk populations require precise and safe diagnostic tools for CAD detection.
Purpose of the Study:
- To prospectively assess the diagnostic accuracy of DVCT for CAD detection in high-risk patients.
- To evaluate the radiation dose associated with DVCT in this population.
- To compare DVCT performance with invasive coronary angiography (ICA).
Main Methods:
- 60 high-risk patients underwent DVCT without heart rate control.
- Invasive coronary angiography (ICA) served as the gold standard for comparison.
- Per-segment analysis was used to determine diagnostic accuracy metrics.
Main Results:
- DVCT demonstrated high diagnostic accuracy for CAD detection (sensitivity 95.3%, specificity 97.6%).
- Diagnostic accuracy was similar across heart rate subgroups, but lower in patients with Agatston score ≥100.
- Mean effective radiation dose was 12.5 ± 9.4 mSv, significantly lower than ICA (14.1 ± 5.9 mSv).
- Radiation dose was significantly lower (5.2 ± 0.9 mSv) when heart rate was <70 bpm compared to >70 bpm (22.6 ± 5.2 mSv).
Conclusions:
- DVCT provides reliable and high diagnostic accuracy for CAD detection, even without heart rate control.
- Heart rate control (<70 bpm) is recommended to minimize radiation exposure during DVCT.
- DVCT is a valuable non-invasive tool for evaluating coronary artery stenosis in high-risk individuals.
Objectives:
The aim of this article was to prospectively evaluate the accuracy and radiation dose of 320-detector row dynamic volume CT (DVCT) for the detection of coronary artery disease (CAD) in a high-risk population.
Methods:
60 patients with a high risk of CAD underwent DVCT without preceding heart rate control and also underwent invasive coronary angiography (ICA), which served as the standard reference.
Results:
On a per segment analysis, overall sensitivity was 95.3%, specificity was 97.6%, positive predictive value was 90.6%, negative predictive value was 98.8% and Youden index was 0.93. In both heart rate subgroups, diagnostic accuracy for the assessment of coronary artery stenosis was similar. The accuracy of the subgroup with an Agatston score ≥100 was lower than that for patients with an Agatston score <100. However, the difference between DVCT and ICA results was not significant (p=0.08). The mean estimated effective dose of CT was 12.5 ± 9.4 mSv. In those patients with heart rates less than 70 beats per minute (bpm), the mean radiation exposure of DVCT was 5.2 ± 0.9 mSv. The effective radiation dose was significantly lower than that of ICA (14.1 ± 5.9 mSv) (p<0.001). When the heart rate was >70 bpm, a significantly higher dose was delivered to patients with DVCT (22.6 ± 5.2 mSv, p<0.001) than with ICA (15.0 ± 5.3 mSv, p<0.001).
Conclusion:
DVCT reliably provides high diagnostic accuracy without heart rate/rhythm control. However, from a dosimetric point of view, it is recommended that heart rate should be controlled to <70 bpm to decrease radiation dose.
More Related Videos
06:57Semi-Automatic Graphical Tool for Measuring Coronary Artery Spatially Weighted Calcium Score from Gated Cardiac Computed Tomography Images
Published on: September 22, 2023
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Acute Coronary Syndrome III: Diagnostic Studies
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Radiological Investigation I: X-ray and CT