Related Experiment Video
Updated: Jun 4, 2026

A Probing Device for Quantitatively Measuring the Mechanical Properties of Soft Tissues during Arthroscopy
Published on: May 1, 2020
Arthrex AR-6400 and AR-6475 arthroscopy pumps can deliver excessive pressure without alerting user
Abstract:
Under certain conditions, Arthrex AR-6400 and AR-6475 arthroscopic irrigation/distention pumps can create pressures within the patient's joint that are more than twice the desired pressure without alerting the clinician, increasing the potential for serious patient injury due to extracapsular solution infiltration. Facilities need to thoroughly understand the behavior of these pumps and determine whether the clinical benefits of using them outweigh the solution-infiltration concerns. If your facility is not comfortable with the behavior of these pumps, use an alternative pump.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
08:12Invasive Hemodynamic Monitoring of Aortic and Pulmonary Artery Hemodynamics in a Large Animal Model of ARDS
Published on: November 26, 2018
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Measurement of Blood Pressure