User experience network. Erroneous downstream occlusion alarms may disable Smiths Medical CADD-Solis infusion pumps
Abstract:
Due to an issue in manufacturing, downstream occlusion (DSO) sensors in some Smiths Medical CADD-Solis infusion pumps may drift out of calibration, potentially resulting in erroneous alarms that disable the units. Hospitals experiencing the problem should return affected units to Smiths Medical for recalibration (free of charge) and should consider testing all their CADD-Solis pumps during routine maintenance to ensure that they alarm appropriately for downstream occlusions.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...