Catheter ablation of right ventricular outflow tract tachycardia: a simplified remote-controlled approach
Melanie Konstantinidou1, Buelent Koektuerk, Erik Wissner
1Department of Cardiology, Asklepios Klinik St Georg, Lohmühlenstrasse 5, Hamburg, Germany. melkondoc@hotmail.com
Aims:
Manual mapping and ablation of right ventricular outflow tract (RVOT) tachycardia may be associated with cardiac perforation and complicated by mechanically induced ventricular extrasystoles (VESs). The aim of this study was to assess remote-controlled magnetic (RCM) mapping and ablation of RVOT ventricular tachycardia (VT)/VES utilizing a soft magnetic catheter.
Methods And Results:
Mapping and ablation of RVOT VT/VES were performed using the magnetic navigation system (MNS, Niobe II, Stereotaxis, St Louis, MO) in conjunction with a cardiodrive motor unit (Stereotaxis). A soft magnetic catheter (Celsius RMT, 4 mm solid tip, Biosense Webster, Diamond Bar, CA) was navigated to the RVOT using a sequence of pre-installed magnetic vectors. The primary endpoint was defined as successful RCM VT/VES ablation. Other parameters observed included RCM RVOT accessibility, mapping-induced mechanical VES, fluoroscopy time, complications, and arrhythmia recurrence. Thirteen consecutive patients (mean age: 50 ± 14 years; 10 females, 6 RVOT VT) underwent RCM mapping and ablation. The RVOT was reached in all patients utilizing solely the MNS. Successful RCM RVOT ablation was achieved in {12 of 13} (92.3%) patients. Mean procedure and fluoroscopy times were 116.5 ± 48.9 and 7.5 ± 4.3 min, respectively. Catheter-induced VES during RVOT mapping were observed in {1 in 13} patients (7.7%, three VES). No complications occurred. During a mean follow-up of 252 ± 211 days, clinical arrhythmia recurrence was observed in {1 in 13} (7.7%) patients.
Conclusions:
Automatic RCM RVOT access is feasible, while RVOT mapping and ablation appear to be safe, fast, and effective. The soft magnetic catheter rarely induces mechanical VES.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
09:13Remote Magnetic Navigation for Accurate, Real-time Catheter Positioning and Ablation in Cardiac Electrophysiology Procedures
Published on: April 21, 2013
Related Concept Videos
Cardiac Catheterization II: Right Heart Catheterization
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias VI: Management of Dysrhythmias
Dysrhythmias III: Characteristics of Dysrhythmias
Cardiac Catheterization III: Left Heart Catheterization