Related Experiment Video
Updated: Jun 4, 2026

A 3D Digital Model for the Diagnosis and Treatment of Pulmonary Nodules
Published on: May 19, 2023
Flat-panel detector computed tomography imaging: observer performance in detecting pulmonary nodules in comparison
Hideji Otani1, Norihisa Nitta, Mitsuru Ikeda
1Department of Radiology, Shiga University of Medical Science, Setatsukinowa-cho, Otsu, Shiga, Japan. otani@belle.shiga-med.ac.jp
Purpose:
The aim of this study was to compare the detectability of lung nodules on images obtained with a flat-panel detector computed tomography (FPD-CT) system and by chest radiographs (CXRs) using receiver-operating characteristic (ROC) analysis.
Materials And Methods:
FPD-CT was conducted with the patients in the sitting position. For the CXR study, the patients stood erect. Our study population consisted of 26 individuals ranging in age from 50 to 83 years. The reference standard was based on the interpretations obtained by consensus of 2 radiologists on multidetector CT images for the presence or absence of nodules. Four other radiologists independently assessed and recorded the absence or presence of lung nodules and their location on FPD-CT and CXR images. ROC analysis was used to evaluate lung nodule detectability by both imaging modalities.
Results:
Two radiologists identified 34 nodules whose diameter was 5 to 42 mm (mean, 19.3 mm) in 23 of the 26 study participants on the multidetector CT images. Overall, analysis of variance for ROC analysis showed that FPD-CT was significantly better in detecting nodules than CXR (P=0.02). The estimated mean Az value was 0.9818±0.0083 with FPD-CT and 0.7610±0.0908 with CXR. The sensitivity for nodule detection on FPD-CT and CXR images was 79.4% and 33.8%, respectively.
Conclusion:
The detectability of pulmonary nodules was better on images of FPD-CT than on CXRs.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies for Cardiovascular System V: CT