Accuracy of working length determination using 3 electronic apex locators and direct digital radiography
Denis Gonçalves Real1, Harry Davidowicz, Cacio Moura-Netto
1School of Dentistry, Universidade Paulista (UNIP), São Paulo, SP, Brazil.
Objectives:
The objectives of this study were to assess the accuracy of working length determination using 3 electronic apex locators and direct digital radiography and to compare the results with those obtained using the visual method (control measurement).
Study Design:
Twenty extracted human maxillary premolars were selected: 17 two-rooted and 3 single-rooted (total of 37 canals). Working length was measured using electronic apex locators Elements Diagnostic, Root ZX, and Just II. Subsequently, teeth were positioned in the alveolar bone of a dry skull and submitted to direct digital radiography. A variation of ±1 mm was considered as acceptable. Results were analyzed using the Wilcoxon and the χ(2) tests.
Results:
Results presented an accuracy of 94.6% for Elements Diagnostic, 91.9% for Root ZX, 73.0% for Just II, and 64.9% for direct digital radiography when considering the margin of ±1 mm in relation to the control measurement. Comparisons with the actual control measurements resulted in accuracy results of 13.51%, 13.51%, 10.10%, and 2.70%, respectively.
Conclusions:
Root ZX and Elements Diagnostic are more accurate in determining working length when compared with Just II and Schick direct digital radiography.