Radiation doses from phantom measurements at high-pitch dual-source computed tomography coronary angiography
Robert Goetti1, Sebastian Leschka, Markus Boschung
1Department of Diagnostic Radiology, University Hospital Zurich, Raemistrasse 100, CH-8091 Zurich, Switzerland. RobertPaul.Goetti@usz.ch
Objective:
To compare radiation doses delivered at prospectively ECG-triggered sequential- (SEQ), retrospectively ECG-gated spiral- (RETRO) and prospectively ECG-triggered high-pitch spiral- (HP) computed tomography coronary angiography (CTCA) protocols, as well as catheter coronary angiography (CCA) using an anthropomorphic phantom.
Materials And Methods:
An anthropomorphic Alderson phantom equipped with 50 thermoluminescent dosimeters (TLDs) was scanned using different CTCA protocols and an uncomplicated diagnostic CCA examination was simulated. Absorbed doses were experimentally determined and effective doses calculated using the dose-length product (DLP) for CTCA and the dose-area product (DAP) for CCA, as well as according to International Commission on Radiation Protection (ICRP) publications 60 and 103.
Results:
Effective organ doses were significantly lower for HP protocols (100kV: 0.17±0.26mSv; 120kV: 0.26±0.39mSv) compared to SEQ protocols (100kV: 0.50±0.79mSv; 120kV: 0.90±1.41mSv; each p<0.05) and compared to RETRO protocols (100kV: 1.59±2.12mSv; 120kV: 2.75±3.50mSv; each p<0.05). Effective organ doses at HP-CTCA tended to be lower than at CCA (0.37±0.40mSv), however this was not statistically significant (p=0.13). Effective doses calculated according to ICRP guidelines could be estimated using the DLP and a conversion coefficient of k=0.034mSv/[mGycm] (ICRP103) or k=0.028mSv/[mGycm] (ICRP60), respectively. HP-CTCA led to a dose reduction of 89% compared to RETRO-CTCA, regardless of the calculation method used.
Conclusions:
Radiation doses as determined by phantom measurements are significantly lower at HP-CTCA compared to SEQ-CTCA and RETRO-CTCA and comparable to uncomplicated diagnostic CCA.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...