Region of interest (ROI) computed tomography (CT): Comparison with full field of view (FFOV) and truncated CT for a
R Chityala1, K R Hoffmann, S Rudin
1Toshiba Stroke Research Center, Rm 445, Biomedical Research Building, State University of New York at Buffalo, 3435, Main Street, Buffalo, NY 14214.
Abstract:
Cone-beam CT reconstruction can be performed at lower integral dose, by using a non-uniform beam filter between the x-ray source and the patient to obtain good image quality within an ROI with minimal artifacts. To evaluate the method, a human head phantom was placed on a rotary stage. Cone-beam projection images of the phantom were obtained with and without an ROI filter (dose reduction factor ~7). A mapping function was established to equalize the intensity outside the ROI (to compensate for the attenuation by the filter) to the intensity inside by assuming that those features lying both inside and outside very close to the edge of the ROI are the same. Reconstructed images were obtained using equalized projection images for 2 cases: one in which the outside region was smoothed using an averaging filter and the other with no smoothing outside. In addition, a third case was simulated by calculating the average pixel value inside the ROI for each image and assigning this value to all pixels outside the ROI for that image. The images were then back projected using a Feldkamp algorithm. We found that the three cases yield results inside the ROI comparable to those obtained using FFOV projections. In addition, the ROI filter reconstruction with smoothing provides image information outside the ROI comparable to the FFOV reconstruction. CT using an ROI filter provides a means to reconstruct reliable 3D for a volume of interest with greatly reduced integral dose compared to FFOV projections and with minimal artifacts.
More Related Videos
08:41Patient-Specific Polyvinyl Alcohol Phantom Fabrication with Ultrasound and X-Ray Contrast for Brain Tumor Surgery Planning
Published on: July 14, 2020
08:02Novel Quantification Protocol for Cardiovascular Calcification Progression Using Longitudinal MicroPET/MicroCT Images
Published on: November 15, 2024
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System V: CT
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...