Evaluation of rebound tonometry in non-human primates
Elizabeth J Elsmo1, Julie A Kiland, Paul L Kaufman
1Dept. of Ophthalmology & Visual Sciences, University of Wisconsin-Madison, Madison, WI 53792, USA.
Experimental Eye Research
|February 15, 2011
Summary
The TonoVet® rebound tonometer accurately measures intraocular pressure (IOP) in cynomolgus macaques. This portable device offers reliable IOP readings, even with varying corneal thickness, making it a valuable tool for primate research.
Area of Science:
- Veterinary Ophthalmology
- Primate Research
- Ophthalmic Instrumentation
Background:
- Intraocular pressure (IOP) measurement is crucial in veterinary ophthalmology, particularly in non-human primates used in research.
- Accurate and reproducible IOP assessment is essential for monitoring ocular health and disease in cynomolgus macaques.
- The TonoVet® rebound tonometer offers potential advantages in ease of use and portability for IOP measurements.
Purpose of the Study:
- To evaluate the accuracy and reproducibility of the TonoVet® rebound tonometer for measuring IOP in anesthetized cynomolgus macaques.
- To investigate the influence of central corneal thickness on TonoVet® IOP measurements in this species.
- To compare TonoVet® readings with manometric IOP and Goldmann applanation tonometry.
Main Methods:
- Anesthetized cynomolgus macaques underwent IOP manipulation via a cannulated anterior chamber connected to a reservoir and transducer.
- IOP was measured using the TonoVet® and 'minified' Goldmann applanation tonometry at various pressure increments and decrements.
- Central corneal thickness was measured using ultrasonic pachymetry; readings were taken by multiple observers.
Main Results:
- TonoVet® readings demonstrated a strong correlation with manometric IOP (r² = 0.955).
- No significant differences in accuracy were found between eyes or observers, though a trend favored right eyes.
- TonoVet® readings during the decremental phase were significantly more accurate than during the incremental phase (p < 0.004).
Conclusions:
- The TonoVet® rebound tonometer is a reliable and accurate instrument for measuring IOP in cynomolgus macaques.
- Its portability, ease of use, and minimal corneal contact make it advantageous, potentially eliminating the need for topical anesthetics.
- The device provides consistent IOP measurements, supporting its use in primate ophthalmic research.