Related Experiment Video
Updated: Jun 4, 2026

Individualized Stem-positioning in Calcar-guided Short-stem Total Hip Arthroplasty
Published on: February 27, 2018
Torsional stability of uncemented femoral stems in oncologic reconstructions
Tom Scharschmidt1, Amy Cohen, Noelle Thomas
1Department of Orthopedics, The Arthur James Cancer Hospital at The Ohio State University, Columbus, OH 43210, USA. thomas.scharschmidt@osumc.edu
Abstract:
Modular oncology implants using uncemented fixation represent a popular reconstruction technique for limb salvage patients. Initial stability is critical to facilitate bony ingrowth of host bone into the stem of a press-fit oncologic modular rotating-hinge total knee arthroplasty (TKA). The impact of stem design on initial stability has not been defined. The goal of this study was to evaluate the initial stability of 3 different stem designs as defined by torsional load to failure. An analysis of imaging was also performed. The pilot study consisted of 5 femora in each of 3 groups based on stem design. The specimen was mounted on a multi-axis biomechanical test frame equipped with a Vicon 3D motion analysis 4-camera system (Vicon Motion Systems, Lake Forest, California) to track the relative motion between the implant and the femur. Torsional force was applied until failure. The straight-fluted stem design had the highest average torsional stiffness (18.3±8.2 Nm/deg) and average torque at 150 μm of implant micromotion (23.2±10.6 Nm) of the 3 stem types tested.The results of this study will help to guide surgical decision making in limb salvage cases. Further investigation is warranted.