Use of a Low-flow Digital Anesthesia System for Mice and Rats
Published on: September 7, 2016
Flushing induction chambers used for rodent anesthesia to reduce waste anesthetic gas
Janet Wolforth1, Melissa C Dyson
1Department of Molecular and Integrative Physiology, the Center for Integrative Genomics, University of Michigan, Ann Arbor, Michigan, USA.
Abstract:
Anesthetic induction chambers used for medical research are a substantial source of waste anesthetic gas (WAG). Ideally, any generated waste gas should be actively vented away from personnel operating the chamber by either a ventilated hood or snorkel. Unfortunately, the ideal environment for anesthetizing rodents is not always available. In an effort to create a safer environment, the authors designed a system to reduce WAG. This system is portable, can be adapted to different precision vaporizing anesthetic systems and fits in a variety of physical locations. The system flushes anesthetic gas out of an induction chamber before operators open the chamber. To ensure that the system was adequately flushing the anesthetic gas, the authors measured WAG concentration in the environment above the induction chamber and directly behind the vent of an activated charcoal filter. They also compared the efficiency of the filters in vertical and horizontal positions. Finally, they measured the recovery time for mice and rats after flushing the anesthetic gas from an induction chamber. The results show that flushing the induction chamber was an inexpensive and effective method for reducing WAG accumulation in the air surrounding the chamber.