Related Experiment Video
Updated: Jun 4, 2026

High Resolution Phonon-assisted Quasi-resonance Fluorescence Spectroscopy
Published on: June 28, 2016
Application of a broadband blue laser diode to trace NO2 detection using off-beam quartz-enhanced photoacoustic
Hongming Yi1, Kun Liu, Weidong Chen
1Key Laboratory of Atmospheric Composition and Optical Radiation, Chinese Academy of Sciences, Hefei, China.
Abstract:
We applied for the first time, to our knowledge, broadband off-beam quartz-enhanced photoacoustic spectroscopy (BB-OB-QEPAS) to trace NO2 detection using a broadband blue laser diode centered at 450 nm. A detection limit of 18 ppbv (parts in 10(9) by volume) for NO2 in N2 at atmospheric pressure was achieved with an average laser power of 7 mW at a 1 s integration time, which corresponds to a 1 σ normalized noise equivalent absorption coefficient of 4.1×10(-9) cm(-1) W=Hz(1=2). An Allan variance analysis was performed to investigate the long-term stability of the BB-OB-QEPAS-based NO2 sensor.