Virtual monochromatic spectral imaging with fast kilovoltage switching: improved image quality as compared with that
Kazuhiro Matsumoto1, Masahiro Jinzaki, Yutaka Tanami
1Department of Diagnostic Radiology and Multi-dimension Biomedical Imaging & Information Laboratory in Research Park, Keio University School of Medicine, 35 Shinanomachi, Shinjuku-ku, Tokyo 160-8582, Japan.
Purpose:
To compare image quality obtained in phantoms with virtual monochromatic spectral (VMS) imaging with that obtained with conventional 120-kVp computed tomography (CT) for a given radiation dose.
Materials And Methods:
Three syringes were filled with a diluted contrast medium (each syringe contained a contrast medium with a different iodine concentration [5, 10, or 15 mg of iodine per milliliter]), and a fourth syringe was filled with water. These syringes were placed in a torso phantom meant to simulate the standard human physique. The phantom was examined with a CT system and use of the fast kilovoltage switching (80 and 140 kVp) and conventional (120 kVp) modes. Image noise and contrast-to-noise (CNR) ratio were analyzed on VMS images and 120-kVp CT images.
Results:
Image noise on VMS images in the range of 67-72 keV was significantly lower than that on the 120-kVp CT images (P < .014). Image noise was lowest at 69 keV and was 12% lower when compared with that on 120-kVp CT images. CNR on the VMS images was highest at 68 keV. CNR on the VMS images obtained at 68 keV in the syringes filled with diluted contrast material (5, 10, and 15 mg of iodine per milliliter) was 28%, 31%, and 30% higher, respectively, compared with that on the 120-kVp CT images (P < .001).
Conclusion:
VMS imaging at approximately 70 keV yielded lower image noise and higher CNR than did 120-kVp CT for a given radiation dose. VMS imaging has the potential to replace 120-kVp CT as the standard CT imaging modality, since optimal VMS imaging may be expected to yield improved image quality in a patient with standard body habitus.
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and solid...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...