Breast mass evaluation: factors influencing the quality of US elastography
Jung Min Chang1, Woo Kyung Moon, Nariya Cho
1Department of Radiology and Clinical Research Institute, Seoul National University Hospital and the Institute of Radiation Medicine, Seoul National University Medical Research Center, 28 Yongon-dong, Chongno-gu, Seoul 100-744, Korea.
Purpose:
To investigate factors influencing the quality of ultrasonographic (US) elastography in the evaluation of suspicious breast masses.
Materials And Methods:
This prospective study was conducted with institutional review board approval; written informed consent was obtained. Between January 2009 and February 2009, real-time US elastography of 312 breast masses (245 benign, 67 malignant) was performed in 268 consecutive patients (mean age, 45.7 years ± 10.2 [standard deviation]) prior to US-guided core biopsy. Five breast radiologists who had performed the examinations assessed the quality of elasticity images as inadequate, low, or high without histologic information. Age, body mass index (BMI), mammographic density, lesion size, lesion depth, and breast thickness at US were analyzed for their association with image quality by using the χ(2) test, Student t test, and multivariate analysis. Sensitivities and specificities for the differentiation of benign from malignant masses on the basis of elastography were calculated and compared between groups of quality scores by using the logistic regression method.
Results:
The quality of elasticity images was assessed as inadequate in 21 (6.7%) cases, low in 134 (42.9%), and high in 157 (50.3%). According to univariate analysis, smaller lesion size (P = .001), shallower lesion depth (P = .005), less breast thickness where the lesion was located (P < .0001), and benign pathologic finding (P = .004) were significantly associated with higher image quality. There was no correlation of image quality with age (P = .213), BMI (P = .191), mammographic density (P = .091), or distance from the nipple (P = .100). Multivariable analysis showed that breast thickness at the location of target lesions was the most important factor influencing elasticity image quality (P = .001). There were significant differences in sensitivity between higher-quality and lower-quality images (87.0% vs 56.8%, respectively; P = .015) in the differentiation of benign from malignant masses.
Conclusion:
Breast thickness at the location of the lesion was the most important factor influencing image quality at US elastography. Sensitivity for classification of benign and malignant masses improved with higher quality scores.
More Related Videos
12:18Magnetic Resonance Elastography Methodology for the Evaluation of Tissue Engineered Construct Growth
Published on: February 9, 2012
05:56Implementation of Non-invasive Point of Care Transient Elastography for Evaluation of Liver Disease in Pediatric Populations with Cystic Fibrosis
Published on: August 29, 2025
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Ultrasonography
During an ultrasonography procedure, a handheld device called a...