Pneumatic performance of the Boussignac CPAP system in healthy humans
Maria Sehlin1, Siv Sandkvist Törnell, Fredrik Öhberg
1Department of Surgical and Perioperative Sciences, Anesthesiology and Intensive Care Medicine, Umeå University, Umeå, Sweden. maria.sehlin@anestesi.umu.se
Background:
The Boussignac continuous positive airway pressure (CPAP) device effectively treats acute pulmonary edema, but data on airway pressure with the Boussignac CPAP system are sparse.
Objective:
To evaluate the Boussignac CPAP system's ability to maintain stable inspiratory and expiratory pressure levels, and to evaluate perceived exertion during breathing with the Boussignac CPAP system.
Methods:
With 18 healthy volunteers we recorded airway pressure and air flow during 10-min sessions at 5.0, 7.5, and 10.0 cm H(2)O. The participants were blinded to the sequence of the CPAP levels. Each session was ended with 10 forced breaths. We measured perceived exertion with the Borg category ratio 10 (Borg CR10) scale.
Results:
When the participants breathed at 20% of vital capacity and a peak expiratory flow of 14% of FEV(1), the maximum pressure difference between inspiration and expiration was 4.0 cm H(2)O at CPAP 10 cm H(2)O. The changes in airway pressure were never large enough to reduce airway pressure to below zero. During the forced breaths, the expiratory volume was 38-42% of vital capacity and peak expiratory flow was 49-56% of FEV(1). As air flow increased, both the drop in inspiratory airway pressure and the increase in expiratory airway pressure increased.
Conclusions:
With CPAP, pressure changes are considered to be associated with increased work of breathing. The device's pneumatic performance is adequate during normal breathing with low air flow, but during forced breathing (high air flow) it did not maintain stable airway pressure, which could increase the work of breathing and cause respiratory fatigue. Thus, the Boussignac CPAP system might be less suitable for a patient breathing at a higher frequency.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Mechanical Ventilation I: Indication and Settings
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Physiological Control of Respiration
Breathing, a seemingly passive process, is regulated by the respiratory center in the brainstem. This center coordinates the involuntary control of respirations, which means it occurs without conscious effort, ensuring a smooth and uninterrupted pattern.
Regulation of Ventilation
The body maintains ventilation by monitoring levels of carbon dioxide (CO2), oxygen (O2), and hydrogen ion concentration (pH) in the arterial blood. Among these factors, the level of CO2 plays a crucial...