Simple Continuous Glucose Monitoring in Freely Moving Mice
Published on: February 24, 2023
Construction of a panel of glucose indicator proteins for continuous glucose monitoring
Sha Jin1, Jithesh V Veetil, Jared R Garrett
1Biomedical Engineering Program, College of Engineering, University of Arkansas, 203 Engineering Hall, Fayetteville, AR 72701, USA.
Abstract:
The development of implantable glucose sensors for use in diabetes treatment has been pursued for decades. However, enzyme-based glucose sensors often fail in vivo. In our previous work, we engineered a novel glucose indicator protein (GIP) that can sense glucose without relying on any enzymes and cofactors. Nevertheless, this GIP is unsuitable for blood glucose monitoring due to its low dissociation constant. Here, we report a novel approach to creating a new GIP that can be used to monitor blood glucose level. By disrupting pi-pi stacking around GIP's glucose binding site through site-directed mutagenesis, we showed that GIP's dissociation constant can be manipulated from 0.026 mM to 7.86 mM. This approach yielded four GIP mutants. We showed that one of the mutants can be used to detect glucose from 0 to 32 mM, while another mutant can be employed to visualize intracellular glucose (0-200 microM) within living cells through FRET imaging microscopy measurement.