Surgical Technique for the Implantation of a Biomimetic Artificial Intervertebral Disc in a Goat Animal Model
Published on: October 10, 2025
The facet joint loading profile of a cervical intervertebral disc replacement incorporating a novel saddle-shaped
Jonathan R Stieber1, Martin Quirno, Mathew Kang
1Department of Orthopaedics, St Luke's Roosevelt Hospital, New York, NY, USA.
Study Design:
Biomechanical study.
Objective:
To evaluate the cervical facet loading profile of an intervertebral cervical disc prosthesis incorporating a unique saddle-shaped articulation and to determine the effect of implantation of the prosthesis on both the loading curve and area distribution of facet loading, when compared with that of an intact specimen.
Summary Of Background Data:
This is the first study that examines the effect of implantation of a cervical disc replacement on the loading of the facet joints.
Methods:
Fresh-frozen ovine cervical spine specimens were used. Specimens were prepared and disarticulated to yield 6 functional spinal units, which were secured and mounted in a custom fixture on a material testing apparatus. A novel pressure sensor was constructed by combining a resistive ink thin film sensor with Fuji pressure-sensitive film. The sensor was calibrated before unilateral insertion into the cervical facet joint via a small arthrotomy. Specimens were tested in both intact condition and after surgical intervention consisting of discectomy and implantation of an intervertebral cervical disc prosthesis (CerviCore, Stryker Spine, Allendale, NJ). Specimens were tested in flexion, extension, and lateral bending to 3 Nm and in torsion to 2.5 Nm. At the termination of each loading profile, the load was maintained for 30 seconds to permit full exposure of the Fuji film. For each configuration, the maximum resultant load, load rate, and contact area pressure at the endpoint of the loading profile were determined. A paired Student t test was used to determine the differences between the intact specimen and the specimen with an implanted intervertebral disc prosthesis.
Results:
There were no statistically significant differences in mean or maximum pressures between the intact specimen and the specimen with an intervertebral disc in all loading configurations. Similarly, there was no significant difference in the total measured force between the groups in all loading configurations. There was no significant difference in contact areas between the groups in flexion, lateral bending, and torsion. When evaluated in extension, the intact specimen had a mean contact area of 0.8 cm compared with 0.5 cm for the intervertebral disc (P<0.02).
Conclusions:
Biomechanical testing in an ovine model demonstrated no significant difference in measured facet pressures and forces between an intact native cervical disc specimen and a cervical intervertebral disc prosthesis using a saddle-shaped articulation. Peak and mean pressures were not demonstrated to be significantly different between the implanted and intact disc conditions. Implantation of the prosthesis resulted in a significant reduction of contact area under extension moments.
Related Concept Videos
Structural Joints: Cartilaginous Joints
There are two types of cartilaginous joints:
Synchondrosis
A synchondrosis ("joined by cartilage") is a cartilaginous joint where bones are connected by hyaline cartilage. Synchondrosis may be temporary or...
Degenerative Disc Disease I: Introduction
Articulations of the Vertebral Column
Degenerative Disc Disease ll: Pathophysiology
Herniated Intervertebral Disc l: Introduction
General Structure of a Vertebra