Assessment of pulp blood flow in primary and permanent teeth using pulse oximetry
Maria Helena Pozzobon1, Ricardo de Sousa Vieira, Ana Maria Hecke Alves
1Department of Endodontics, School of Dentistry, Universidade Federal de Santa Catarina Department of Pediatric Dentistry, School of Dentistry, Universidade Federal de Santa Catarina, Florianópolis, Santa Catarina, Brazil. lelenapozzobon@yahoo.com.br
Abstract:
Pulse oximetry (PO) is a well-accepted non-invasive method for assessing vascular health, based on oxygen saturation (SaO(2) ) measurements. The objectives of this study were to design and build a custom-made PO dental sensor holder, to evaluate the effectiveness of PO in determining pulp blood flow in primary and permanent teeth, and to compare the SaO(2) levels obtained in teeth and on the little finger of patients. The PO sensor adapted to the custom-made sensor holder is termed as a device to position and hold the PO sensor. This study evaluated SaO(2) readings obtained in the pulp of 123 teeth of 84 children aged 4-13years. The teeth were divided into three groups: group I - primary teeth: 23 central incisors and 28 canines; group II - permanent teeth: 25 central incisors and 28 canines; and group III - endodontically treated teeth (control): 12 permanent central incisors and seven primary central incisors. The pulp blood flow and SaO(2) were measured and were compared with readings of the patient's finger. Data were analyzed by paired t-tests and Pearson's analysis method. The PO was able to identify all the clinically normal pulps contained in the sample, and all the endodontically treated teeth (controls) showed no response. The mean SaO(2) values were 85.27% in the teeth and 92.85% in the fingers. In conclusion, PO readings were effective in determining pulp blood flow in primary and permanent teeth. However, there was no correlation between the SaO(2) values in the fingers and in the teeth of the patients (P<0.05).
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for:
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Special considerations while measuring pulse