Related Experiment Video
Updated: Jun 4, 2026

Three-Dimensional Echocardiographic Method for the Visualization and Assessment of Specific Parameters of the Pulmonary Veins
Published on: October 28, 2020
Pulmonary vein inflow artifact reduction for free-breathing left atrium late gadolinium enhancement
Mehdi H Moghari1, Dana C Peters, Jouke Smink
1Department of Medicine, Cardiovascular Division, Harvard Medical School, Beth Israel Deaconess Medical Center, Boston, Massachusetts 02215, USA.
Abstract:
Two-dimensional "pencil-beam" navigator, placed on the right hemidiaphragm, is used for free-breathing late gadolinium enhancement of the left atrium in patients with atrial fibrillation. The pencil-beam navigator creates an inflow artifact in the right pulmonary veins and atrial wall that may obscure local pulmonary vein and left atrium scars. To reduce this artifact, we propose a large slab right hemidiaphragm projection navigator that measures the respiratory motion while reducing the associated inflow artifact. Eighteen subjects underwent pulmonary vein late gadolinium enhancement using the pencil-beam and projection navigator. Subjective inflow and respiratory motion artifact scores (1 = severe, 2 = moderate, 3 = mild, and 4 = none) from two blinded readers were compared. The artifact scores were 3.8 ± 0.4 and 2.1 ± 0.7 for the projection and pencil-beam navigators, respectively (P < 0.001). Respiratory motion artifact scores were similar between the two techniques (3.0 ± 0.5 vs. 3.1 ± 0.5 for projection vs. pencil-beam navigator). The proposed method greatly reduces the inflow artifact in free-breathing pulmonary vein late gadolinium enhancement while allowing adequate respiratory motion compensation.