4D dose-position verification in radiation therapy using the RADPOS system in a deformable lung phantom
Amanda Cherpak1, Monica Serban, Jan Seuntjens
1Medical Physics, The Ottawa Hospital Cancer Centre, Ottawa, Ontario K1H 8L6, Canada. acherpak@ottawahospital.on.ca
Medical Physics
|March 3, 2011
Summary
The RADPOS system, used with a deformable lung phantom, shows promise as a quality assurance tool for 4D radiotherapy. It accurately measures dose and position during simulated treatments, aiding in adaptive radiotherapy delivery.
Area of Science:
- Medical Physics
- Radiation Oncology
- Quality Assurance
Background:
- 4D radiotherapy requires precise monitoring of tumor motion during treatment.
- Current quality assurance methods for 4D radiotherapy may not fully capture dynamic treatment scenarios.
Purpose of the Study:
- To evaluate a novel 4D in vivo dosimetry system (RADPOS) as a quality assurance tool for 4D radiotherapy.
- To assess the accuracy of RADPOS in measuring position and dose within a deformable lung phantom.
Main Methods:
- RADPOS detectors (MOSFET dosimeter + electromagnetic positioning probe) were placed in a deformable lung phantom.
- Positional accuracy was assessed by comparing RADPOS readings to CT data.
- Dose measurements were performed during simulated free-breathing and adaptive radiotherapy delivery scenarios.
Main Results:
- RADPOS demonstrated positional accuracy better than 2.5 mm, with radial displacement agreement within 0.7 mm between CT and linac rooms.
- Dose measurements during free-breathing delivery agreed within 4-5% of treatment planning doses.
- Adaptive radiotherapy delivery showed RADPOS dose values within 1.5% of treatment plan values.
Conclusions:
- The deformable lung phantom-RADPOS system is an efficient quality assurance tool for 4D radiotherapy.
- The system's accuracy in position and dose measurement supports its use in advanced radiotherapy techniques.


