Related Experiment Video
Updated: Jun 3, 2026

Detection of Fluorescent Nanoparticle Interactions with Primary Immune Cell Subpopulations by Flow Cytometry
Published on: March 28, 2014
The use of flow cytometry for the detection of subvisible particles in therapeutic protein formulations
Henryk Mach1, Akhilesh Bhambhani, Brian K Meyer
1Merck Research Laboratories, Bioprocess Analytical and Formulation Sciences, West Point, Pennsylvania 19486, USA. henryk.mach@merck.com
Abstract:
The amount, identity, and size distribution of particles in parenteral therapeutic protein formulations are of immense interest due to potential safety and efficacy-related implications. In this communication, we describe the use of a flow cytometer equipped with forward- and side-scattering as well as fluorescence detectors, to determine the number of subvisible particles in monoclonal antibody formulations. The method appears to detect particles of size 1 μ and larger, requiring relatively small sample volumes to estimate subvisible particle counts. Additionally, it facilitates differentiation of proteinaceous particles after staining with a fluorescent hydrophobic dye. The method is expected to be particularly well suited for pharmaceutical development, because it provides increased throughput due to the use of a 96-well autosampler.

