Related Experiment Video
Updated: Jun 3, 2026

06:38
Establishment of a Segmental Femoral Critical-size Defect Model in Mice Stabilized by Plate Osteosynthesis
Published on: October 12, 2016
Mechanical analysis of resorbable plates for long-term soft-tissue molding
Davood Varghai1, Harvey Chim, Arun K Gosain
1Department of Plastic Surgery, Case Western Reserve University, University Hospitals of Cleveland, Cleveland, Ohio 44106, USA.
The Journal of Craniofacial Surgery
|March 16, 2011
Summary
Resorbable plates used in pediatric craniofacial surgery lack the mechanical rigidity for long-term soft-tissue molding. Further development is needed to improve their stability in prolonged out-of-plane bends.
Area of Science:
- Biomaterials Science
- Orthopedic Surgery
- Pediatric Craniofacial Surgery
Background:
- Resorbable plates are increasingly utilized in pediatric craniofacial surgery.
- Novel applications involve long-term soft-tissue molding, though plates weren't designed for this.
- The mechanical stability of these plates under prolonged out-of-plane bending remains unaddressed.
Purpose of the Study:
- To evaluate the mechanical rigidity and stability of resorbable plates during prolonged out-of-plane bending.
- To assess the suitability of current resorbable plate designs for soft-tissue molding applications.
Main Methods:
- Ten types of resorbable plates were bent to a right angle.
- Plates were subjected to an 80-g load at body temperature for 2 weeks.
- Angle height was measured to assess maintenance of out-of-plane bend, with 3 successive tests per plate.
Main Results:
- All tested resorbable plates exhibited a decrease in angle height over the 2-week period.
- Plate thickness correlated with mechanical rigidity and the ability to sustain out-of-plane bends.
- Synthes PLGA (85/15) 1.2-mm and BioGeneral PLA (poly-D,L-lactic acid) 1.0-mm plates demonstrated superior performance.
Conclusions:
- No resorbable plates tested maintained their initial out-of-plane bend over 2 weeks.
- Current resorbable plates are inadequate for long-term soft-tissue molding applications requiring sustained out-of-plane stability.
- Industry collaboration is necessary to enhance resorbable plate mechanical properties for these specialized uses.