Accuracy and precision evaluation of seven self-monitoring blood glucose systems
Chih-Yi Kuo1, Cheng-Teng Hsu, Cheng-Shiao Ho
1Department of Clinical Laboratory, Tai-An Hospital, Taichung, Taiwan.
Background:
Self-monitoring blood glucose (SMBG) systems play a critical role in management of diabetes. SMBG systems should at least meet the minimal requirement of the World Health Organization's ISO 15197:2003. For tight glycemic control, a tighter accuracy requirement is needed.
Methods:
Seven SMBG systems were evaluated for accuracy and precision: Bionime Rightest(™) GM550 (Bionime Corp., Dali City, Taiwan), Accu-Chek(®) Performa (Roche Diagnostics, Indianapolis, IN), OneTouch(®) Ultra(®)2 (LifeScan Inc., Milpitas, CA), MediSense(®) Optium(™) Xceed (Abbott Diabetes Care Inc., Alameda, CA), Medisafe (TERUMO Corp., Tokyo, Japan), Fora(®) TD4227 (Taidac Technology Corp., Wugu Township, Taiwan), and Ascensia Contour(®) (Bayer HealthCare LLC, Mishawaka, IN). The 107 participants (44 men and 63 women) were between 23 and 91 years old. The analytical results of seven SMBG systems were compared with those of plasma analyzed with the hexokinase method (Olympus AU640, Olympus America Inc., Center Valley, PA).
Results:
The imprecision of the seven blood glucose meters ranged from 1.1% to 4.7%. Three of the seven blood glucose meters (42.9%) fulfilled the minimum accuracy criteria of ISO 15197:2003. The mean absolute relative error value for each blood glucose meter was calculated and ranged from 6.5% to 12.0%.
Conclusions:
More than 40% of evaluated SMBG systems meet the minimal accuracy criteria requirement of ISO 15197:2003. However, considering tighter criteria for accuracy of ±15%, only the Bionime Rightest GM550 meets this requirement. Because SMBG systems play a critical role in management of diabetes, manufacturers have to strive to improve accuracy and precision and to ensure the good quality of blood glucose meters and test strips.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Measurement of Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.