End-expired nitrous oxide concentrations compared to flowmeter settings during operative dental treatment in children
Ulrich Klein1, T Jay Robinson, Amanda Allshouse
1Department of Pediatric Dentistry, The Children's Hospital and School of Dental Medicine, University of Colorado Denver, Aurora, Colo, USA. klein.ulrich@tchden.org
Purpose:
The purpose of this study was to analyze inspired and end-expired alveolar nitrous oxide (N(2)O) concentrations during operative dental treatment in 6- to 9- year-old children and compare them to the concentrations dispensed by the dental N(2)O delivery system's flowmeter.
Methods:
Twenty-three healthy 6- to 9-year-olds who received restorative dental care underwent placement of a sampling probe in the posterior nasopharynx. Via nasal mask, 100% oxygen was administered at a rate of 3L/minute for 2 minutes. N(2)O was then introduced at a flowmeter reading of 50% for 5 minutes and thereafter reduced to 30% for the duration of treatment, followed by 100% oxygen for 5 minutes postoperatively. Respiratory rate, inspired and end-expired nitrous oxide and oxygen concentrations, and pressure of end-tidal carbon dioxide values were measured and recorded from the nasopharyngeal probe with an anesthesia gas monitor.
Results:
End-expired alveolar N(2)O concentrations were, on average, 63% below flowmeter settings. After rapid induction with 50% N(2)O, an end-expired N(2)O saturation maximum of 11% was attained, on average, after 90 seconds. The time to flush out remaining N(2)O after delivery of 100% oxygen varied between 30 and 195 seconds.
Conclusions:
End-expired alveolar nitrous oxide and oxygen concentrations and pressure of end-tidal carbon dioxide values were successfully recorded during operative dental treatment of 6- to 9-year-old awake and non-intubated children in the traditional dental setting.
More Related Videos
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Inhalational Anesthetics: Overview