Reliability and concurrent validity of the perometer for measuring hand volume in women with and without lymphedema
Mi-Joung Lee1, Robert A Boland, Sharon Czerniec
1Discipline of Physiotherapy, Faculty of Health Sciences, University of Sydney, Sydney, Australia. mi-joung.lee@sydney.edu.au
Background:
Lymphedema of hand after breast cancer treatment causes significant loss of hand function. Although there are several ways of assessing limb volume, measuring hand volume has been problematic due to technical difficulties associated with assessment of finger volumes. The aim of this study was to investigate the criterion validity and reliability of Perometer™ for measuring hand volume in woman with and without lymphedema.
Methods And Results:
Hand volume of forty women with (n = 20) and without lymphedema (n = 20) was assessed twice by one rater and once by another rater using the Perometer, and once by one rater using the water volumetry method. Intra- and inter-rater reliability was determined from the intraclass correlation coefficients and Percent Close Agreement. Agreement between the Perometer and water volumetry was determined using a limit of agreement and Lin's concordance correlation. The Perometer had high intra [ICC(2,1) = 0.989 (95% CI: 0.98-0.99)] and inter-rater reliability [ICC(2,1) = 0.993 (95% CI: 0.99-1.0)]. Percent close agreement revealed that 80% of the measures were within 9 ml for inter-rater reliability and within 15 ml for intra-rater reliability. In addition, there was high concordance between hand volumes obtained with the Perometer and water volumetry method (R(c) = 0.88). However, the Perometer overestimated the volume of hand compared to water volumetry method (bias: 7.5%).
Conclusions:
The Perometer can be used with high reliability to measure hand volume but caution should be exercised when data are compared with measures derived from the water volumetry method.
Related Concept Videos
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above the...
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of the Abdomen II: Percussion
Percussion
Percussion is an essential...
Assessment of blood pressure in brachial artery(two-step method)
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement