Assessment of trending ability of cardiac output monitors by polar plot methodology
Lester A Critchley1, Xiao X Yang, Anna Lee
1Department of Anaesthesia & Intensive Care, The Chinese University of Hong Kong, Prince of Wales Hospital, Shatin, New Territories, Hong Kong, China. hcritchley@cuhk.edu.hk
Objectives:
To develop a valid statistical method of showing acceptable cardiac output (CO) trending ability when new CO monitors are compared to a reference standard, such as thermodilution, using polar coordinates.
Design:
Developing a new statistical analytic method using historic data.
Setting:
University Hospital Anesthesia and Intensive Care Department.
Participants:
Data taken from previously published CO validation studies.
Interventions:
Cartesian data were reanalyzed, being uplifted using Data Thief 3.0 software (http://datathief.org/). Polar plots were constructed from this data. Central zone data (<0.5 L/min or <10% change) were excluded because they introduced statistical noise. Trial polar criteria were set using data from a study that compared 5 CO monitors against thermodilution. Then, these criteria were further validated using data extracted from 15 other studies. Mean (95% confidence intervals) polar angles were used.
Measurements And Main Results:
Trial data suggest ±5° (angle) ±30° (95% confidence interval) as acceptance limits. Concordance rates (ie, >95%-90%) from 5 articles supported trending, and polar data from these studies concurred with the authors' pilot criteria. Favorable comments on trending also were found in 8 of 15 articles in which radial limits were less than ±32°. Good calibration was associated with a mean polar angle of less than ±5°.
Conclusions:
Polar plots can be used to show the trending ability of CO monitors in comparative validation studies. They overcome the deficiencies of concordance analysis, which uses the direction of change as a statistic and ignores the magnitude of change in CO.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send blood...