Reproducibility of the new Goldmann AT900D digital tonometer
Laura Morales-Fernandez1, Jose M Martinez-de-la-Casa, Julian Garcia-Feijoo
1Departamento de Oftalmologia, Hospital Clinico San Carlos, Instituto de Investigaciones Oftalmologicas Ramon Castroviejo, Universidad Complutense, Madrid, Spain.
Purpose:
To examine the reproducibility of the new digital Goldmann applanation tonometer AT 900D (DGT) for measuring intraocular pressure and to compare its readings with those provided by the conventional analog Goldmann tonometer (AGT).
Methods:
In 3 different sessions (S1, S2, and S3), 3 intraocular pressure measurements were made in 15 eyes of 15 healthy subjects using both the DGT and AGT. The order of use of the tonometers was random and the data were collected by a different examiner to the one taking the measurements. Data were compared by determining coefficients of variation for each session and representing the differences detected as Bland-Altman plots.
Results:
The mean differences recorded (AGT-DGT) were 0.1, 0.3, and -0.1 for S1, S2, and S3, respectively. None of the differences were significant and none were ≥3 mm Hg. Coefficients of variation were lower in the 3 sessions for the DGT, the difference being significant for S2 and S3 (4.87% vs. 3.99%, P=0.400; 4.84% vs. 3.55%, P=0.028; and 5.04% vs. 3.50%, P=0.025 in S1, S2, and S3 respectively).
Conclusions:
The new digital Goldmann tonometer shows a high reproducibility and excellent correlation of measures with those of the AGT.
Related Concept Videos
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Uncertainty in Measurement: Reading Instruments
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure