Clinical Anthropometrics and Body Composition from 3-Dimensional Optical Imaging
Published on: June 7, 2024
[Challenges for computed tomography of overweight patients]
F Bamberg1, R Marcus, M Petersilka
1Institut für Klinische Radiologie, Klinikum der Ludwig-Maximilians-Universität München, Campus Grosshadern, München, Deutschland.
Dual source XXL mode computed tomography (CT) provides diagnostic image quality for morbidly obese patients weighing over 150 kg, unlike standard CT protocols. While XXL mode increases radiation dose, it is essential for reliable imaging in this patient group.
Area of Science:
- Medical Imaging
- Radiology
- Diagnostic Imaging
Background:
- Morbidly obese patients often require computed tomography (CT) for non-invasive diagnostics.
- Standard CT scanners have limitations in weight capacity, dose, and image quality for this population.
Purpose of the Study:
- To analyze weight limits, radiation dose, and image quality of standard CT scanners.
- To determine the diagnostic value and radiation dose of a dual source XXL mode CT scanner for morbidly obese patients.
Main Methods:
- Retrospective analysis of 15 morbidly obese patients (average weight 189.6 ± 42 kg) examined with the XXL mode.
- Intraindividual comparison of XXL and standard protocols in 7 patients.
- Inclusion of 14 patients (90-150 kg) examined with standard protocols as references.
- Examination of a large abdomen phantom to compare dose efficiency.
Main Results:
- The dual source XXL mode resulted in significantly higher radiation doses (CTDIvol XXL 72.9 ± 23 mGy vs. standard 16.7 ± 11 mGy).
- Image noise was higher in the XXL group but significantly lower in intraindividual comparisons (e.g., liver: 24.2 ± 14 HU vs. 36.3 ± 20 HU).
- The XXL protocol achieved diagnostic image quality in all cases, whereas 43% of standard protocol examinations were non-diagnostic.
Conclusions:
- Standard CT techniques are suitable for patients up to 150 kg.
- The dual source XXL mode is necessary for reliable diagnostic image quality in patients exceeding 150 kg, despite increased radiation dose.
- Standard dose conversion factors may overestimate the biologically relevant equivalent dose in XXL mode scans.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...