Computerized Dynamic Posturography for Postural Control Assessment in Patients with Intermittent Claudication
Published on: December 11, 2013
[Peripheral arterial disease: efficacy of the oscillometric method]
Jorge Vega1, Sebastián Romaní, Francisco J Garcipérez
1Servicio de Cardiología, Hospital San Pedro de Alcántara, Cáceres, Spain. jorgevegafernandez@gmail.com
Abstract:
Relatively little is known on how the Doppler method compares with oscillometric measurement using a conventional automatic blood pressure device to determine the ankle-brachial index, when determinations are performed by physicians with little experience. To assess the diagnostic efficacy of both methods in this professional group, we calculated their sensitivity, specificity, and positive and negative predictive value in 158 legs of 85 patients with symptoms of intermittent claudication. Angiography was used as the gold standard. Of the legs examined, 131 showed significant arterial obstruction. The oscillometric method showed 97% sensitivity, 89% specificity, 98% positive predictive value, and 86% negative predictive value. The Doppler method showed 95% sensitivity, 56% specificity, 91% positive predictive value, and 68% negative predictive value. This study suggests that the automatic blood pressure equipment has greater diagnostic accuracy when the test is performed by physicians not specifically trained to use the Doppler probe. Full English text available from: www.revespcardiol.org.
Related Concept Videos
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of blood pressure in brachial artery(two-step method)