Evaluation of a Point-of-Care Testing Analyzer for Measuring Peripheral Blood Leukocytes
Published on: March 22, 2022
An automated image analysis system can be beneficial in preclassification of leucocytes in children with
Oytun Portakal1, Betul Tavil, Barış Kuşkonmaz
1Clinical Pathology Laboratory, Hacettepe University Medical School, Ankara, Turkey. oytun@hacettepe.edu.tr
Abstract:
This study was aimed to evaluate the analytical performance of an automated image analysis system (a pilot model of Diff Master(™) Octavia) for the preclassification of leucocytes in children with hematological disease. Manual microscopy performed by pediatric hematologists was used as the reference method. Five mature cell class and blasts were evaluated. Diff Master Octavia correctly preclassified 87.4% of all leucocytes with a high reproducibility. The overall accuracy was found to be 93.0%. Clinical sensitivity was 97.7% and specificity was 76.0%. The average time per slide for Diff Master(™) Octavia was 2.3 min lower than that of manual method. Our results indicated that the Diff Master(™) Octavia can detect and preclassify leucocytes accurately; therefore, it can be used as an efficient and fast method in pediatric hematology routine.