Purpose:
Inadequate airway humidification can result in endotracheal tube occlusion. There is evidence that heat and moisture exchangers (HMEs) are more prone to endotracheal tube occlusion than heated humidifiers (HHs) that contain a heated wire circuit. We aimed to compare the incidence of endotracheal tube occlusion while introducing a new dual-heated wire circuit HH in place of an established hydrophobic HME.
Materials And Methods:
This was a prospective observational study. All patients who required intubation were included in our analysis. Univariate statistical analysis was performed using a Fisher exact test. P < .05 was considered statistically significant.
Results:
There were 158 patients in the HME group and 88 patients in the HH group. The incidence of endotracheal tube occlusion was 5.7% in the HME group and 0% in the HH group. Statistical analysis revealed a significant difference between the 2 groups (P = .02). In light of this finding, we changed our practice to provide humidification exclusively by HH. In the subsequent 18-month period, there were no further episodes of endotracheal tube occlusion.
Conclusions:
Our study demonstrates that there is a significant increase in the incidence of endotracheal tube occlusion when using a hydrophobic HME compared with an HH and that using a dual-heated wire circuit HH can eliminate endotracheal tube occlusion.
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...