Improving image quality in portal venography with spectral CT imaging
Li-qin Zhao1, Wen He, Jian-ying Li
1Department of Radiology, Beijing Friendship Hospital Affiliated to Capital Medical University, Beijing,100050, China. zhaolqzr@sohu.com
Objective:
To investigate the effect of energy spectral CT on the image quality of CT portal venography in cirrhosis patients.
Materials And Methods:
30 portal hypertension patients underwent spectral CT examination using a single-tube, fast dual tube voltage switching technique. 101 sets of monochromatic images were generated from 40keV to 140keV. Image noise and contrast-to-noise ratio (CNR) for portal veins from the monochromatic images were measured. An optimal monochromatic image set was selected for obtaining the best CNR for portal veins. The image noise and CNR of the intra-hepatic portal vein and extra-hepatic main stem at the selected monochromatic level were compared with those from the conventional polychromatic images. Image quality was also assessed and compared.
Results:
The monochromatic images at 51keV were found to provide the best CNR for both the intra-hepatic and extra-hepatic portal veins. At this energy level, the monochromatic images had about 100% higher CNR than the polychromatic images with a moderate 30% noise increase. The qualitative image quality assessment was also statistically higher with monochromatic images at 51keV.
Conclusion:
Monochromatic images at 51keV for CT portal venography could improve CNR for displaying hepatic portal veins and improve the overall image quality.
More Related Videos
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Portal Hypertension
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies VII: Vascular Imaging
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies