Some practical comparisons of the efficiency and overloading behaviour of sub-2 μm porous and sub-3 μm shell
1Centre for Research in Biomedicine, University of the West of England, Frenchay, Bristol, UK. David.Mccalley@uwe.ac.uk
Abstract:
At their optimum flow, sub-3 μm superficially porous or "shell" particles demonstrate similar efficiency to sub-2 μm totally porous particles. The performance of 0.21 cm i.d shell columns is however inferior to those of 0.46 cm i.d., presumably due to packing difficulties. At high flow, shell columns can give flatter Knox curves due to lower operating pressure (half or less of that of the totally porous particles) producing less frictional heating, which combined with the increased thermal conductivity of their non-porous core, gives more efficient heat dissipation. However, the effects of frictional heating for sub-2 μm columns are considerably exaggerated when using pure ACN as mobile phase, as it has a thermal conductivity 3 times less than that of pure water, leading to poorer heat dissipation. Overloading is already problematic for ionised solutes, a group which contains many pharmaceuticals and compounds of clinical relevance, on conventional columns (5 μm porous particles). However, it becomes a more serious issue for both new column types, partially as a result of their very high efficiency, which concentrates the sample as a very narrow band. The sample capacity of one type of shell particle was estimated to be 60% of that of the small totally porous particles, in line with the fraction of the particle volume that is porous. Due to overloading, it is barely possible to achieve perfect peak symmetry for ionised acids or bases with either of these new column types, even by injecting the lowest amounts of sample detectable by UV. While ammonium formate and potassium phosphate buffers gave similar results in overloading studies, use of formic acid as sole mobile phase additive is not recommended for these solutes, as its ionic strength is too low, leading to a catastrophic deterioration in efficiency when sample concentrations of even a few mg/L are injected.
Related Concept Videos
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Size-Exclusion Chromatography
Silica particles offer advantages such as rigidity,...
Principles Of Column Chromatography