Related Experiment Video
Updated: Jun 3, 2026

A Method for Quantifying Upper Limb Performance in Daily Life Using Accelerometers
Published on: April 21, 2017
Detection of seizure-like movements using a wrist accelerometer
Juliana Lockman1, Robert S Fisher, Donald M Olson
1Department of Neurology and Neurological Sciences, Stanford University Medical Center, Stanford, CA 94304-1510, USA.
A new wrist-worn motion detector can effectively detect tonic-clonic seizures in epilepsy patients. This wearable device alerts caregivers to seizures, potentially preventing injuries and improving safety for individuals with epilepsy.
Area of Science:
- Neurology
- Biomedical Engineering
- Medical Devices
Background:
- Caregivers of epilepsy patients often worry about injuries or death from unwitnessed seizures.
- Tonic-clonic seizures pose significant risks due to their unpredictable nature.
Purpose of the Study:
- To evaluate the efficacy of a wrist-worn motion detector in identifying tonic-clonic seizures.
- To assess the device's potential for alerting caregivers to seizure events.
Main Methods:
- A wrist-worn motion detector was used in individuals undergoing continuous video/electroencephalogram (EEG) monitoring.
- The device detected rhythmic movements characteristic of tonic-clonic seizures and transmitted alerts via Bluetooth.
- Device detections were cross-referenced with video/EEG recordings.
Main Results:
- The motion detector successfully detected 7 out of 8 tonic-clonic seizures in 6 out of 40 patients.
- The device recorded 204 non-seizure movements, with a patient-cancellable alert system.
- Only one false detection occurred during sleep, indicating high specificity.
Conclusions:
- A wrist-worn motion detector shows promise for detecting tonic-clonic seizures.
- This technology could provide timely alerts to caregivers, enhancing patient safety and monitoring.
More Related Videos
06:28High-Quality Seizure-Like Activity from Acute Brain Slices Using a Complementary Metal-Oxide-Semiconductor High-Density Microelectrode Array System
Published on: September 27, 2024
10:25Multi-system Monitoring for Identification of Seizures, Arrhythmias and Apnea in Conscious Restrained Rabbits
Published on: March 27, 2021