A novel technique for stenting pulmonary artery and conduit bifurcation stenosis
Oliver Stumper1, Vinay Bhole, Ben Anderson
1Heart Unit, Birmingham Children's Hospital, United Kingdom. oliver.stumper@bch.nhs.uk
Background:
Distal conduit obstruction is a recognized complication after surgery for congenital heart disease requiring implantation of a conduit from the right ventricle to the pulmonary arteries. Endovascular stenting of distal conduit obstruction can be challenging due to the proximity to the pulmonary artery bifurcation.
Objective:
A technique is described, whereby a single stent is mounted onto two balloon angioplasty catheters in tandem. This ensemble was delivered to the distal conduit/pulmonary artery via a large Mullins sheath on two guidewires, one placed in each of the branch pulmonary arteries. The aim was to assess safety and efficacy of this novel technique.
Materials And Results:
Seven patients (mean age 13.4 (6.7-23.4) years, mean weight 44.2 (23-69) kg were treated with this method. The pressure gradient was reduced from 36 (26-52) mm Hg to 11 (8-15) mm Hg [P< 0.05]. RV/LV pressure ratio decreased from 0.85 (0.6-0.95) to 0.42 (0.35-0.5) [P < 0.05]. There were no significant complications. During follow-up over a median of 2.6 (0.3-6.7) years no patient required re-intervention or surgery.
Conclusion:
This novel technique appears to be safe and effective for stenting stenoses just proximal to pulmonary artery bifurcation.