Related Experiment Video
Updated: Jun 3, 2026

Utilizing vmTracking to Improve the Accuracy of Multi-Animal Pose Estimation in Rodent Social Behavior Studies
Published on: November 7, 2025
Performance evaluation of real-time motion tracking using positron emission fiducial markers
Marc Chamberland1, Richard Wassenaar, Benjamin Spencer
1Department of Physics, Carleton University, 1125 Colonel By Drive, Ottawa, Ontario K1S 5B6, Canada.
Purpose:
Tumor motion due to patient breathing is a factor that limits the accuracy of dose distribution in radiotherapy. One of the methods to improve the accuracy is by applying respiratory gating or tumor tracking. Both techniques require a precise determination of the spatial location of the tumor. We present an experimental evaluation of the performance of PeTrack, a technique that can track internal fiducial markers in real-time for tumor tracking.
Methods:
PeTrack uses position sensitive detectors to record annihilation coincidence gamma rays from fiducial positron emission markers implanted in or around the tumor. It uses an expectation-maximization clustering algorithm to track the position of the markers. A normalized least mean square adaptive filter was used to predict the position of the markers 100 and 200 ms in the future. We evaluated the performance of the tracking and of the prediction by using a dynamic anthropomorphic thorax phantom to generate three-dimensional (3D) motion of three fiducial markers. The algorithm was run with four different data sets. In the first run, the motion of the markers was based on a sinusoidal model of respiratory motion. Three additional runs were done with motion based on patient breathing data.
Results:
In the case of the sinusoidal model, the average 3D root mean square error for all markers was 0.44 mm. For the three runs based on patient breathing data, the precision of the 3D localization was 0.49 mm. At a latency of 100 ms, the average 3D prediction error was 1.3 +/- 0.6 mm for the sinusoidal model and for the three patient breathing runs. At a latency of 200 ms, the average 3D prediction errors were 1.7 +/- 0.8 mm for the sinusoidal model and 1.4 +/- 0.7 mm for the breathing runs.
Conclusions:
We conclude that PeTrack can track multiple fiducial markers in real-time with an accuracy and precision smaller than 2 mm. PeTrack can have a direct application in tumor tracking for radiation therapy.